Answers are grounded in this paper and cite their sources.
The Cs-137 irradiator is widely used to irradiate biological samples for radiobiological research. To obtain the accurate outcomes, correct measurements of the delivered absorbed dose to a sample is important. The IAEA protocols such as TRS-277 and TRS-398 were recommended for the Cs-137 reference dosimetry. However in TRS-398 protocol, currently known as the most practical dosimetry protocol, the quality factor ( ) for Cs-137 gamma rays is not suggested. Therefore, the use of TRS-398 protocol is currently unavailable for the Cs-137 dosimetry directly. The calculation method previously introduced for high energy photon beams in radiotherapy was used for deriving the Cs-137 beam qualities ( ) for the 15 commercially available farmer type ionization chambers in this study. In conclusion, values were ranged from 0.998 to 1.002 for Cs-137 gamma rays.
No takes yet. Share an insight, caveat, or question.
Kang et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: