Answers are grounded in this paper and cite their sources.
Experience with a number of deep X-ray therapy units has shown that variations in X-ray output occur, which are often due to changes in the kilo-voltage operating across the X-ray tube. By utilising the well-known property that, over a range of quality, suitably chosen filters of different atomic number behave differently as regards their X-ray absorption, a simple method for detecting such changes in potential has been developed, which can be used conveniently in an X-ray therapy department. The method consists simply in measuring the intensity of the X-ray beam when (a) heavily and (b) lightly filtered. The ratio of the two intensities is found to serve as a sensitive indication of the applied kilovoltage, and hence of the quality of the beam. The filters usually met with in an X-ray therapy department serve well for the purpose of the test.
No takes yet. Share an insight, caveat, or question.
D. E. A. Jones (1939) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: