Answers are grounded in this paper and cite their sources.
In order to evaluate a new method of heparinization, uncoated (22) and heparin-coated (27) central venous polyethylene catheters were inserted in 49 patients via basilic and cephalic veins punctured at the fossa cubiti. The means duration of catheterization was 5.7 (1-11) days. One-third of the patients with heparin-coated catheters, and one sixth with uncoated catheters developed clinical thrombophlebitis, with a maximum incidence between 4 and 8 days after catheterization. A higher risk of developing thrombophlebitis in the first 4 days after catheterization was found in the patients with heparin-coated polyethylene catheters. After 8 days of catheterization, it seems that there is a lower risk of new cases of thrombophlebitis appearing both in patients with uncoated and those with heparin-coated polyethylene catheters. Radiological thrombosis, regardless of duration of catheterization and heparin-coating, was demonstrated in all 22 patients investigated by "pull-out" phlebography. The heparin-coating did not decrease the rate of thrombotic complications. Location of the catheter tip in subclavian veins was associated with a significantly higher incidence of large, parietal thrombi and catheter occlusion than when the tip was situated in anonymous veins, the superior vena cava, or the right atrium. Cannulation by heparin-coated, polyethylene tubing did not reduce the rate of catheter occlusion.
No takes yet. Share an insight, caveat, or question.
Bennegård et al. (1982) studied this question.