Why the study?
Does impedance cardiography provide reproducible measurements of cardiac output and other haemodynamic parameters during postural changes and exercise?
Does impedance cardiography provide reproducible measurements of cardiac output and other haemodynamic parameters during postural changes and exercise?
Non-invasive impedance cardiography provides a reproducible means to evaluate cardiac output and other hemodynamic variables during different postures and light-to-moderate exercise.
Answers are grounded in this paper and cite their sources.
Supports impedance cardiography for serial hemodynamic assessment across conditions; extends validation data but leaves open effects on clinical outcomes.
Background: Evaluation of cardiac output (CO) and other haemodynamic parameters may aid in understanding factors involved in arterial blood pressure (BP) changes with exercise and postural stress.Impedance cardiography offers a rapid, non-invasive means to acquire this information, however there is limited data assessing the reproducibility of this technique during haemodynamic perturbation.This study aimed to assess reproducibility of CO and other haemodynamic parameters derived from impedance cardiography during exercise and in different postures.Methods: 51 participants (mean age 57 AE 9 years, 57% male) had CO and other haemodynamic variables (including end diastolic volume, left ventricular work, ejection fraction and systemic vascular resistance) measured via impedance cardiography (Physio Flow) at two visits separated by 12 AE 7 days.Measures were recorded at rest in three postures (supine, seated and standing), during upright cycle ergometry at a fixed workload (40 W), and also during steady state exercise at an intensity of 60% and 70% of age-predicted maximum heart rate (HR max ).Results: CO reproducibility was assessed over a wide range (5.27 AE 1.00e12.09AE 2.02 l/min).There was good agreement between CO measured at each visit in all postures and exercise conditions (intra-class correlation coefficient [ICC] range 0.729e0.888,P < 0.05 for all) with a small difference between visits (mean difference 0.06 AE 1.10 l/min).All other haemodynamic variables showed good agreement between visits (ICC range 0.714e0.970,P < 0.05 for all).Conclusions: Non-invasive impedance cardiography provides an acceptably reproducible means to evaluate CO and other haemodynamic variables relevant to arterial BP regulation during different postures and light-to-moderate intensity exercise.
No takes yet. Share an insight, caveat, or question.
Schultz et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: