Answers are grounded in this paper and cite their sources.
Changes in ambient pressure affects the reflectivity of ultrasound contrast microbubbles leading to an excellent correlation between subharmonic signals and hydrostatic pressure. The aortas of two dogs were scanned with an experimental pulse-echo system to validate in vivo pressure estimation based on subharmonic microbubble signals. Results matched well with instantaneous pressure measurements (from 20-60 mmHg) obtained simultaneously with a pressure catheter (root mean square errors <27%).
No takes yet. Share an insight, caveat, or question.
Forsberg et al. (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: