Why the study?
Interchangeability of IVC measurements between subcostal and trans-hepatic windows is not fully established, and automated border detection using AI requires validation.
Does AI-based automated border detection provide accurate IVC distensibility measurements compared to standard M-mode in mechanically ventilated patients?
Does AI-based automated border detection provide accurate IVC distensibility measurements compared to standard M-mode in mechanically ventilated patients?
AI software provides good accuracy but suboptimal precision with wide limits of agreement compared to M-mode for assessing IVC distensibility in mechanically ventilated patients.
AI may facilitate IVC assessment when subcostal view fails; leaves open whether trans-hepatic measurements guide fluid decisions.
Background: Variation of inferior vena cava (IVC) is used to predict fluid-responsiveness, but the IVC visualization with standard sagittal approach (SC, subcostal) cannot be always achieved. In such cases, coronal trans-hepatic (TH) window may offer an alternative, but the interchangeability of IVC measurements in SC and TH is not fully established. Further, artificial intelligence (AI) with automated border detection may be of clinical value but it needs validation. Methods: Prospective observational validation study in mechanically ventilated patients with pressure-controlled mode. Primary outcome was the IVC distensibility (IVC-DI) in SC and TH imaging, with measurements taken both in M-Mode or with AI software. We calculated mean bias, limits of agreement (LoA), and intra-class correlation (ICC) coefficient. Results: Thirty-three patients were included. Feasibility rate was 87.9% and 81.8% for SC and TH visualization, respectively. Comparing imaging from the same anatomical site acquired with different modalities (M-Mode vs AI), we found the following IVC-DI differences: 1)SC: mean bias -3.1%, LoA [-20.1;13.9], ICC=0.65; 2)TH: mean bias -2.0%, LoA [-19.3;15.4], ICC=0.65. When comparing the results obtained from the same modality but from different sites (SC vs TH), IVC-DI differences were: 3)M-Mode: mean bias 1.1%, LoA [-6.9;9.1], ICC=0.54; 4)AI: mean bias 2.0%, LoA [-25.7;29.7], ICC=0.32. Conclusions: In patients mechanically ventilated, AI software shows good accuracy (modest overestimation) and moderate correlation as compared to M-mode assessment of IVC-DI, both for SC and TH windows. However, precision seems suboptimal with wide LoA. The comparison of M-Mode or AI between different sites yields similar results but with weaker correlation.
No takes yet. Share an insight, caveat, or question.
Via et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: