Why the study?
Does noninvasive ultrasound Doppler accurately determine the pressure gradient compared to manometry in patients with a Björk-Shiley mitral valve?
Does noninvasive ultrasound Doppler accurately determine the pressure gradient compared to manometry in patients with a Björk-Shiley mitral valve?
Ultrasound Doppler is an accurate and reliable noninvasive method for determining the mean diastolic pressure gradient in patients with a Björk-Shiley mitral valve implant.
Answers are grounded in this paper and cite their sources.
May support Doppler for Björk-Shiley mitral gradients; leaves open prospective validation before practice change.
The accuracy in determining the pressure gradient in the Björk-Shiley mitral implant from noninvasive ultrasound Doppler data was explored in nine adult patients. Manometric and ultrasound data were collectd simultaneously, and identical diastolic periods were used to compare the manometric gradient (delta Pm) with the gradient obtained from ultrasound data (delta Pu). In the nine patients the mean diastolic value of delta Pm ranged from 2-12.5 mm Hg and the difference between the mean diastolic values of delta Pm and delta Pu was 0.3 +/- 1.0 mm Hg(SD). The results of the investigation indicated that the method is accurate and reliable in the nonivasive determination of the mean diastolic gradient in the Björk-Shiley mitral implant.
No takes yet. Share an insight, caveat, or question.
Holen et al. (1979) studied this question.