Why the study?
Cardiac T1ρ dispersion quantification for sensitive detection of myocardial fibrosis without contrast agents is challenging because multi-amplitude T1ρ mapping is very time-consuming.
A novel Bloch simulation-optimized radial sampling pattern enables fast and accurate high-resolution myocardial T1ρ mapping and dispersion quantification in mice.
Answers are grounded in this paper and cite their sources.
Enables feasible non-contrast T1ρ dispersion mapping for fibrosis detection; leaves open prospective clinical validation.
Purpose T 1ρ dispersion quantification can potentially be used as a cardiac magnetic resonance index for sensitive detection of myocardial fibrosis without the need of contrast agents. However, dispersion quantification is still a major challenge, because T 1ρ mapping for different spin lock amplitudes is a very time consuming process. This study aims to develop a fast and accurate T 1ρ mapping sequence, which paves the way to cardiac T 1ρ dispersion quantification within the limited measurement time of an in vivo study in small animals. Methods A radial spin lock sequence was developed using a Bloch simulation-optimized sampling pattern and a view-sharing method for image reconstruction. For validation, phantom measurements with a conventional sampling pattern and a gold standard sequence were compared to examine T 1ρ quantification accuracy. The in vivo validation of T 1ρ mapping was performed in N = 10 mice and in a reproduction study in a single animal, in which ten maps were acquired in direct succession. Finally, the feasibility of myocardial dispersion quantification was tested in one animal. Results The Bloch simulation-based sampling shows considerably higher image quality as well as improved T 1ρ quantification accuracy (+ 56%) and precision (+ 49%) compared to conventional sampling. Compared to the gold standard sequence, a mean deviation of − 0.46 ± 1.84% was observed. The in vivo measurements proved high reproducibility of myocardial T 1ρ mapping. The mean T 1ρ in the left ventricle was 39.5 ± 1.2 ms for different animals and the maximum deviation was 2.1% in the successive measurements. The myocardial T 1ρ dispersion slope, which was measured for the first time in one animal, could be determined to be 4.76 ± 0.23 ms/kHz. Conclusion This new and fast T 1ρ quantification technique enables high-resolution myocardial T 1ρ mapping and even dispersion quantification within the limited time of an in vivo study and could, therefore, be a reliable tool for improved tissue characterization.
No takes yet. Share an insight, caveat, or question.
Gram et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: