Are defibrillation energy requirements stable between implant and 3-month follow-up in patients with a nonthoracotomy lead ICD system?
Are defibrillation energy requirements stable between implant and 3-month follow-up in patients with a nonthoracotomy lead ICD system?
A new weighted logistic regression analysis demonstrated that defibrillation energy requirements for a nonthoracotomy lead ICD system remain stable from implant to 3-month follow-up.
Answers are grounded in this paper and cite their sources.
Defibrillation thresholds appear stable from implant to 3-month follow-up in nonthoracotomy ICDs; extends prior observations but should not yet change practice.
BACKGROUND: A weighted logistic regression analysis was developed to allow pooling of patient data for the study of the stability of defibrillation energy requirements with a new nonthoracotomy lead defibrillation system. METHODS AND RESULTS: One hundred twenty patients were prospectively studied with a single-model nonthoracotomy implantable cardioverter defibrillator (ICD) system at the time of implant and at 3 months. The pooled data of all shocks delivered to all patients were fitted to a logistic function to construct a defibrillation voltage/energy dose-response relationship. The crude logit curve was weighted in quartiles according to the average shock energy delivered per patient. Shocks at implant (n = 802; 6.6 +/- 2.5 shocks/patient) and follow-up (n = 292; 2.4 +/- 1.2 shocks/patient) were analyzed. The modeled voltage/energy required for 50% successful defibrillation (95% CI) in the pooled data was 367 V (273, 461) and 9.8 J (6.7, 12.9) at implant and 338 V (264, 412) and 10.5 J (8, 13.0) at follow-up. The conventional measure of lowest successful voltage/energy (95% CI) was 430 V (411, 449) and 12.1 J (11, 13.2) at implant and 415 V (391, 439) and 11.3 J (10, 12.6) at follow-up. There were no statistically significant differences between implant and follow-up energy requirements with either method. CONCLUSIONS: The nonthoracotomy lead system used in this study demonstrated stability of defibrillation energy requirements at implant and 3-month follow-up. A new technique for the estimation of the defibrillation energy dose-response relationship was derived by using a weighted logistic regression analysis.
No takes yet. Share an insight, caveat, or question.
Newman et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: