Why the study?
Does a defibrillation safety margin of 10 J improve first shock success rate compared to a safety margin equal to the defibrillation energy requirement in patients with low defibrillation energy requirements?
Does a defibrillation safety margin of 10 J improve first shock success rate compared to a safety margin equal to the defibrillation energy requirement in patients with low defibrillation energy requirements?
In patients with a low defibrillation energy requirement (≤ 6 J), a safety margin of 10 J provides a higher rate of successful defibrillation than a safety margin equal to the defibrillation energy requirement.
Answers are grounded in this paper and cite their sources.
May support a 10 J safety margin in low-DEF patients; hypothesis-generating and requires randomized confirmation before practice change.
INTRODUCTION: In patients undergoing defibrillator implantation, an appropriate defibrillation safety margin has been considered to be either 10 J or an energy equal to the defibrillation energy requirement. However, a previous clinical report suggested that a larger safety margin may be required in patients with a low defibrillation energy requirement. Therefore, the purpose of this prospective study was to compare the defibrillation efficacy of the two safety margin techniques in patients with a low defibrillation energy requirement. METHODS AND RESULTS: Sixty patients who underwent implantation of a defibrillator and who had a low defibrillation energy requirement (< or = 6 J) underwent six separate inductions of ventricular fibrillation, at least 5 minutes apart. For each of the first three inductions of ventricular fibrillation, the first two shocks were equal to either the defibrillation energy requirement plus 10 J (14.6+/-1.0 J), or to twice the defibrillation energy requirement (9.9+/-2.3 J). The alternate technique was used for the subsequent three inductions of ventricular fibrillation. For each induction of ventricular fibrillation, the first shock success rate was 99.5%+/-4.3% for shocks using the defibrillation energy requirement plus 10 J, compared to 95.0%+/-17.2% for shocks at twice the defibrillation energy requirement (P = 0.02). The charge time (P < 0.0001) and the total duration of ventricular fibrillation (P < 0.0001) were each approximately 1 second longer with the defibrillation energy requirement plus 10 J technique. CONCLUSION: This study is the first to compare prospectively the defibrillation efficacy of two defibrillation safety margins. In patients with a defibrillation energy requirement < or = 6 J, a higher rate of successful defibrillation is achieved with a safety margin of 10 J than with a safety margin equal to the defibrillation energy requirement.
No takes yet. Share an insight, caveat, or question.
Strickberger et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: