Does the orientation of a tricuspid bioprosthetic mitral valve influence the flow field inside the left ventricle in an in vitro model?
Does the orientation of a tricuspid bioprosthetic mitral valve influence the flow field inside the left ventricle in an in vitro model?
In vitro modeling demonstrates that the orientation of a mitral bioprosthesis significantly alters intraventricular flow fields and aortic upstream velocity profiles.
Answers are grounded in this paper and cite their sources.
Valve orientation may alter intraventricular flow; in vitro PIV data leave optimal mitral bioprosthesis positioning open.
The flow patterns of a prosthetic heart valve in the aortic or mitral position can change according to its type and orientation. This work describes the use of 2D particle image velocimetry (PIV) applied to the in vitro flow fields characterization inside the upper part of a left ventricular model at various heart rates and as a function of two orientations of stented tricuspid mitral bioprostheses. In the ventricular model, each mitral bioprosthesis (27 and 31 mm diameter) was installed in two orientations, rotated by 180°, while the aortic bileaflet mechanical valve (27 mm diameter) remained in a fixed orientation. The results (N = 50) showed changes in the intraventricular flow fields according to the mitral bioprostheses positioning. Also, changes in the aortic upstream velocity profiles were noticed as a function of mitral orientations.