Why the study?
It was unclear how aortic flow measurements with increasing distance from the aortic valve and breath-hold regimen affect calculated mitral regurgitation assessed by the indirect cardiac MRI method.
Does the location of aortic flow measurement in cardiac MRI affect the calculation of mitral regurgitation by the indirect method in patients without structural abnormalities?
Does the location of aortic flow measurement in cardiac MRI affect the calculation of mitral regurgitation by the indirect method in patients without structural abnormalities?
Cardiac MRI using the indirect method systematically overestimates mitral regurgitation, and this overestimation worsens as flow measurements are taken further from the aortic valve.
Answers are grounded in this paper and cite their sources.
Distal aortic flow sites may overestimate MR by indirect CMR; supports standardized proximal planes in future studies.
OBJECTIVE: Cardiac MRI is quickly emerging as the gold standard for assessment of mitral regurgitation, most commonly with the indirect method subtracting forward flow in aorta from volumetric segmentation of the left ventricle. We aimed to investigate how aortic flow measurements with increasing distance from the aortic valve affect calculated mitral regurgitations and whether measurements were influenced by breath-hold regimen. METHODS: Free-breathing and breath-hold phase contrast flows were measured in aorta at valve level, sinotubular (ST) junction, mid-ascending aorta and in the pulmonary trunk. Flow measurements were pairwise compared, and subsequently, after exclusion of patients with visible mitral and tricuspid regurgitations for left-sided and right-sided comparisons, respectively, flow-measured stroke volumes were compared with ventricular volumetric segmentations. RESULTS: Thirty-nine participants without arrhythmias or structural abnormalities of the large vessels were included. Stroke volumes measured with free-breathing and breath-hold flow decreased equally with increasing distance to the aortic valves (breath-hold flow: aortic valve 105.6±20.8 mL, ST junction 101.5±20.7 mL, mid-ascending aorta 98.1±21.5 mL). After exclusion of atrioventricular regurgitations, stroke volumes determined by volumetric measurements were higher compared with values determined by flow measurements, corresponding to 'false' atrioventricular regurgitations of 8.0%±5.8% with flow measured at valve level, 11.6%±5.2% at the ST junction and 15.3%±5.0% at the mid-ascending aorta. CONCLUSIONS: Stroke volumes determined by flow decrease throughout the proximal aorta and are systematically lower than volumetrically measured stroke volumes. The indirect method systematically overestimates mitral regurgitations, especially with increasing distance from the aortic valves.
No takes yet. Share an insight, caveat, or question.
Bertelsen et al. (2020) studied this question.