November 1, 1996Pacing and Clinical Electrophysiology

Effect of Varying Atrial Sensitivity, AV Interval, and Detection Algorithm on Automatic Mode Switching

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Do varying atrial sensitivity, AV interval, and detection algorithms affect the efficacy of automatic mode switching in patients with pacemakers?

Population

18 patients with pacemakers having automatic mode switching capability who developed atrial fibrillation or…

Comparison

Programming changes to pacemaker settings… vs Baseline pacemaker settings.

Design

Case_series

Authors

EPEugen PalmaAlbert Einstein College of MedicineVKVAIDYA KEDARNATHMontefiore Medical CenterVVVIREN VANKAWALLAMontefiore Medical Center

Discussion

Loading...

Member takes

Overview

May warrant individualized AMS programming checks in AF; Level 5 case report leaves open prospective validation.

Structured PICO

Do varying atrial sensitivity, AV interval, and detection algorithms affect the efficacy of automatic mode switching in patients with pacemakers?

P
Population
18 patients (6 women; mean age 64 years) with pacemakers having automatic mode switching (AMS) capability who developed atrial fibrillation (n=14) or atrial flutter (n=4).
I
Intervention
Programming changes to pacemaker settings including varying atrial sensitivity, increasing AV intervals from 60 to 200 ms, and making detection algorithms more stringent (from 5 beats at 150 beats/min to 11 beats at 200 beats/min).
C
Comparator
Baseline pacemaker settings (normal sinus rhythm sensitivity, shorter AV intervals, less stringent detection algorithms).
O
Outcome
Efficacy of automatic mode switching (AMS) during atrial tachyarrhythmia.surrogate

Effective automatic mode switching in pacemakers during atrial fibrillation requires higher atrial sensitivity settings, shorter AV intervals, and less stringent detection algorithms.

Cite This Study

Palma et al. (1996) studied this question.

synapsesocial.com/papers/6a71be4526a7f98052ddf3cbhttps://doi.org/10.1111/j.1540-8159.1996.tb03216.x
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1DDD Pacing: An Effective Treatment Modality for Recurrent Atrial Arrhythmias1988 · 108 citations
  2. 2Morbidity and Mortality of Patients with Sinus Node Disease: Comparative Effects of Atrial and Ventricular Pacing1990 · 58 citations
  3. 3Chronic atrial fibrillation and stroke in paced patients with sick sinus syndrome. Relevance of clinical characteristics and pacing modalities.1993 · 205 citations
  4. 4Amplitude of Atrial Electrical Activity During Sinus Rhythm and During Atrial Flutter‐Fibrillation1985 · 24 citations
  5. 5A New Feature for Control of Inappropriate High Rate Tracking in DDDR Pacemakers1990 · 26 citations