BACKGROUND: Transpulmonary thermodilution allows the measurement of cardiac index for high risk surgical patients. Oncologic patients often have a central venous access (port-a-catheter) for chronic treatment. The validity of the measurement by a port-a-catheter of the absolute cardiac index and the detection of changes in cardiac index induced by fluid challenge are unknown. METHODS: We conducted a monocentric prospective study. 27 patients were enrolled. 250 ml colloid volume expansions for fluid challenge were performed during ovarian cytoreductive surgery. The volume expansion-induced changes in cardiac index measured by transpulmonary thermodilution by a central venous access (CIcvc) and by a port-a-catheter (CIport) were recorded. RESULTS: 23 patients were analyzed with 123 pairs of measurements. Using a Bland and Altman for repeated measurements, the bias (lower and upper limits of agreement) between CIport and CIcvc was 0.14 (-0.59 to 0.88) L/min/m2. The percentage error was 22%. The concordance between the changes in CIport and CIcvc observed during volume expansion was 92% with an r = 0.7 (with exclusion zone). No complications (included sepsis) were observed during the follow up period. CONCLUSIONS: The transpulmonary thermodilution by a port-a-catheter is reliable for absolute values estimation of cardiac index and for measurement of the variation after fluid challenge. TRIAL REGISTRATION: clinicaltrials.gov NCT02063009.