Key result
The fully automated bioluminescent enzyme immunoassay (BLEIA) demonstrated a good correlation with real-time quantitative RT-PCR for detecting norovirus in fecal samples.
The fully automated BLEIA provides rapid and sensitive detection of norovirus, comparable to molecular methods, with a 46-minute turnaround time.
Answers are grounded in this paper and cite their sources.
May support rapid norovirus testing in labs; leaves open prospective validation of clinical impact.
An ultrasensitive and fully automated bioluminescent enzyme immunoassay (BLEIA) was developed for the detection of norovirus (NV) capsid antigen. In the evaluation tests with recombinant virus-like particles, the BLEIA demonstrated broad reactivity against several NV genotypes (genotypes 1, 3, 4, 7, 8, and 12 in genogroup I [GI] and genotypes 1, 2, 3, 4, 5, 6, 12, and 13 in GII), a wide dose-response range from 0.25 pg/ml to 10,000 pg/ml, and good reproducibility with low coefficients of variation (CVs) (within-run CVs of <2.8%, between-day CVs of <3.7%). In the evaluation tests with NV-positive fecal samples, a good correlation (y = 0.66x - 3.21, r = 0.84) between the BLEIA and real-time quantitative reverse transcription-PCR was obtained. Furthermore, in the dilution test with NV specimens, the analytical sensitivity of NV was estimated to be 10(5) to 10(6) copies/g of fecal sample, indicating that the analytical sensitivity of the BLEIA is comparable to that of commercially available molecular methods. All assay steps are fully automated, the turnaround time is 46 min, and the throughput of the assay is 120 tests/h. These results indicate that the BLEIA is potentially useful for the rapid diagnosis of NV in epidemic and sporadic gastroenteritis.
No takes yet. Share an insight, caveat, or question.
Sakamaki et al. (2012) studied Norovirus infection. Bioluminescent enzyme immunoassay (BLEIA) vs. Real-time quantitative reverse transcription-PCR was evaluated on Correlation between BLEIA and real-time quantitative reverse transcription-PCR (r = 0.84). The fully automated bioluminescent enzyme immunoassay (BLEIA) demonstrated a good correlation with real-time quantitative RT-PCR for detecting norovirus in fecal samples.
Synapse has enriched 10 related papers. Consider them for comparative context: