This case demonstrates the feasibility of transapical mitral paravalvular leak closure and reports the first successful percutaneous apical closure of a large bore (14 F) sheath using a vascular plug device.
Answers are grounded in this paper and cite their sources.
Highlights risks after repeated MVR in rheumatic disease; leaves open optimal long-term strategies.
A 55-year-old female presented to our institution for recurrent hospitalization due to dyspnea and pulmonary edema. The patient is known to have a history of rheumatic heart disease status post-mitral valve replacement (MVR) 3 times before. The first MVR was at the age of 18 using a bovine valve. The first re-do was done 15 years later, at the age of 33, using a mechanical valve. At the age of 51, the patient underwent both aortic valve replacement (AVR) and second re-do MVR due to valve thrombosis using a mechanical bi-leaflet valve. This was complicated with severe medial mitral para-valvular regurgitation that was treated with a vascular plug a couple of years back. Repeat trans-esophageal echo (TEE) revealed severe medial paravalvular leak adjacent to the previous plug directed toward the interatrial septum (Video 1 and 2). A mitral paravalvular leak (PVL) closure was planned due to the prohibitive risk for a third redo-MVR. Due to the medial location of the defect with a thick and fibrotic septum, a transapical approach was chosen as it provides direct access to the mitral valve, allows easy access to defects and increases the likelihood of successful mitral PVL closure with less radiation exposure. The procedure was performed under general anesthesia. The apex was punctured using ultrasound guidance with a micro-puncture needle (Video 3). A 6Fr sheath was then introduced over a J wire. The medial mitral PVL defect was then crossed with a Terumo wire and 5 F Judkins right 4 catheter (Video 4). A 7Fr shuttle delivery sheath was then introduced in the defect over a stiff wire placed in the left atrium. The defect measured 7 × 4 mm on 3D TEE (Figure 1a) and an Occulotech PLD Occluder 8 × 4 mm (Figure 1b) was implanted under TEE guidance with no residual leak (Video 5). However, after deployment, restriction of one of the mechanical valve leaflets was observed (Video 6). Repositioning of the plug with a snare resulted in embolization of the plug in the left ventricle and was flipping back and forth between the two mechanical valves (Video 7 and 8). Attempt to snare the device in a retro-aortic fashion was unsuccessful as the device was getting stuck in the mechanical valve. A Bi-directional snaring from the apex and retro-aorta was used to elongate the device and attempt to remove across the mechanical aortic valve was also unsuccessful. Eventually, we were able to snare the device after upsizing the apical access to a large 14 F sheath (Video 9 and 10). An AVP 3 6 × 3 mm was then used to close the defect successfully (Video 11), with complete resolution of the leak seen by TEE (Video 12). The apical access was closed using an AMPLATZER Vascular Plug II 18 mm with evidence of good hemostasis after protamine administration (Video 13 and 14). Post deployment, the patient had increased peaked pressure on the ventilator and was found to have hemothorax that was managed with a chest tube (Figure 2a,b). The patient was discharged 4 days later and is now 6 months post-PVL closure with complete resolution of symptoms with no heart failure readmissions, New York Heart Association (NYHA) class I symptoms, with TTE showing no PVL and apical plug in place (Video 15) and significant decrease in PA pressure. To our knowledge, this is the first successful reported percutaneous apical closure with a vascular plug device for a large bore sheath (14 F). None.
No takes yet. Share an insight, caveat, or question.
Zgheib et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: