Why the study?
Heart rate signals obtained from micro-Doppler radar for noncontact vital sign monitoring are subject to noise interference caused by imperfect circuit characteristics.
A 120-GHz millimeter-wave radar using intelligent SVD and improved AMPD algorithms achieves high accuracy (96.53%-98.18%) for noncontact short-term heart rate measurement compared to pulse wave measurements.
May improve noncontact radar heart rate accuracy; leaves open clinical validation in cardiovascular patients.
Heart rate measurement based on radar is an effective method for noncontact monitoring of human vital signs. However, the heart rate signal obtained from the radar is subject to noise interference due to imperfect circuit characteristics. To tackle this challenge in heart rate detection by micro-Doppler radar, this article uses a narrow-beam radar operating in the 120-GHz band to capture the heartbeat signal and adopts an intelligent singular value decomposition (SVD) algorithm to effectively eliminate noise in the heartbeat signal. This approach can adaptively select the number of large singular values corresponding to the best noise reduction effect under different signal-to-noise ratios (SNRs). In addition, an improved automatic multiscale peak detection (AMPD) algorithm is proposed for precise peak detection in the heartbeat signal. Experimental results demonstrate that the proposed algorithm significantly reduces omission or misjudgment during the peak detection process, leading to a high accuracy during short-term heart rate detection. Compared with the pulse wave measurement results, the average accuracy of the heart rate is between 96.53% and 98.18%.
No takes yet. Share an insight, caveat, or question.
Lv et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: