Why the study?
Heartbeat signals from Doppler radar sensors lack clear peak values unlike ECGs or pulse waves, limiting measurement accuracy.
Does a 120 GHz narrow beam FMCW millimeter wave radar system accurately measure short-term heart rate compared to PPG in volunteers?
Does a 120 GHz narrow beam FMCW millimeter wave radar system accurately measure short-term heart rate compared to PPG in volunteers?
A novel 120 GHz narrow beam FMCW millimeter wave radar system can accurately measure short-term heart rate without contact, showing high consistency with standard PPG.
May enable comfortable continuous monitoring; leaves open large-scale clinical validation before adoption.
Heart rate measurement technology based on Doppler radar can monitor heart rate in real time without direct contact with the skin, which possesses the advantages of convenient operation, comfort and insensitivity to the skin state. However, due to the limitation of radar’s performance, heartbeat signals obtained by radar sensor do not have clear peak value like electrocardiograms (ECGs) and pulse waves, which affects its measurement accuracy. Therefore, this paper proposes a non-contact short-term heart rate detection system based on 120 GHz narrow beam frequency-modulated continuous-wave (FMCW) millimeter wave radar. On the basis of improving the signal-to-clutter ratio (SCR) of intermediate frequency (IF) signals with narrow beam, this system applies stochastic resonance algorithm to enhance the amplitude of heartbeat signals submerged in radar system noise, thus enhancing the signal-to-noise ratio (SNR) of heartbeat signals and improving the accuracy of heart rate detection. The short-term heart rate of 10 volunteers is measured at 3 s interval under the condition that the radar sensor was 1 m from the human chest. Experimental results show that compared with the peak count of the reference pulse wave sensor, the results of this system are highly consistent with those of photoplethysmograph (PPG) detection. Finally, those results of heart rate detection are also used to analyze heart rate variability (HRV).
No takes yet. Share an insight, caveat, or question.
Lv et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: