Why the study?
The gold-standard conventional FD-TWA device was commercially discontinued, and normal reference values for a newly available 24-hour AECG-based FD-TWA system were lacking.
This study establishes normal reference values for a novel 24-hour ambulatory ECG-based frequency domain T-wave alternans measurement in subjects without structural heart disease.
Answers are grounded in this paper and cite their sources.
Provides initial normal values for 24-h FD-TWA via new AECG; leaves open validation for arrhythmic risk stratification.
BACKGROUND: Conventional frequency domain T wave alternans (FD-TWA) is a noninvasive risk stratification marker for identifying arrhythmic sudden cardiac death, but the conventional FD-TWA device that was considered the gold standard device has been discontinued commercially. Recently, a newly developed ambulatory electrocardiogram (AECG) device that can detect FD-TWA continuously for 24 hours is available in clinical settings. However, information on the normal values using the novel AECG-based frequency domain TWA (FD-TWA) is lacking. METHODS: FD-TWA for AECG was examined in 312 subjects without heart disease (Sb-wHD) (range 20-89 years, 146 men) and 30 heart disease patients (HD-P) (mean age 57±17 years, 24 men). The maximum FD-TWA amplitude over 24 hours was measured with manual editing. The upper limit of local noise levels for measurement of FD-TWA was set to both <10 μV and <20 μV (acceptable noise level <10 μV and <20 μV). RESULTS: The reference values (95th percentiles) of FD-TWA in Sb-wHD were 19.9 μV for the acceptable noise level <10 μV and 23.6 μV for the acceptable noise level <20 μV. The 75th percentile of FD-TWA amplitude in HD-P was 19.5 µV at an acceptable noise level <10 µV and 21.5 µV at an acceptable noise level <20 µV. FD-TWA amplitude without heart disease was significantly affected by heart rate when the maximum FD-TWA was measured (β = 0.274 p < 0.001 for the acceptable noise level <10 μV; β = 0.263, p < 0.001 for the acceptable noise level <20 μV) and age (β = 0.204, p = 0.004 for the acceptable noise level <10 μV; β = 0.149, p = 0.034 for the acceptable noise level <20 μV). CONCLUSIONS: In the present study, the reference values for a novel FD-TWA in Sb-wHD and the distribution of TWA values in HD-P were established. In future research, the cut-off values of FD-TWA in HD-P will need to be examined.
No takes yet. Share an insight, caveat, or question.
Hashimoto et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: