Why the study?
Does frame and volume rate affect the concordance between 2D, 3D, and triplane speckle tracking strain in patients evaluated for subclinical LV function?
Does frame and volume rate affect the concordance between 2D, 3D, and triplane speckle tracking strain in patients evaluated for subclinical LV function?
Strain values are not interchangeable among 2D, 3D, and triplane speckle tracking, with poor correlations driven by differences in acquisition rates, making 2D strain currently the most robust measurement.
Answers are grounded in this paper and cite their sources.
Strain values from different modalities should not be interchanged in subclinical LV assessment; this cohort study leaves open whether volume rate optimization enhances 3D concordance.
OBJECTIVE: To assess the effects of frame and volume rate on the concordance between two-dimensional speckle tracking strain (2DS) and three-dimensional speckle tracking strain (3DS), and between 2DS and triplane imaging of speckle tracking (Tri-P). METHODS: Global longitudinal strains (GLSs) derived from 2DS, 3DS, and Tri-P were compared among 142 prospectively recruited patients who underwent evaluation of subclinical left ventricle (LV) function. RESULTS: Feasibility to obtain GLS of 3DS was significantly higher than that of Tri-P (76% vs. 47%, P < 0.001). The correlation between 2DS and 3DS was only modest (r = 0.47) whereas that of 2DS and Tri-P was better (r = 0.67). The difference in frame/volume rate between two methods also affected their correlation. A volume rate between 34 and 50 volumes/sec had the highest correlation between 2DS and 3DS (r = 0.72). The correlation between 2DS and Tri-P was better with a difference in frame-rate ≤20 per second than with a difference >20 per second. Likewise, there was a better correlation between 2DS and 3DS when the difference between 2D frame rate and 3D volume rate was ≤40 per second, compared to when it was >40 per second. These associations differed from segment to segment and the apical segments had the highest correlation and the basal the lowest. CONCLUSIONS: The feasibility of each means of strain calculation showed important differences, with 2DS being the most attainable. Strain values were not interchangeable among 2DS, Tri-P, and 3DS. Importantly, poor correlations seemed to be driven by differences in acquisition rate. Currently, 2DS offers the most robust measurement of subclinical myocardial dysfunction.
No takes yet. Share an insight, caveat, or question.
Negishi et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: