Why the study?
ECG contamination impairs diaphragm electromyography analysis, and standard manual removal is time-consuming and prone to inter-rater disagreement.
Does a semi-automated analysis method improve reliability and reduce analysis time compared to manual analysis of diaphragm electromyography in patients with COPD?
Does a semi-automated analysis method improve reliability and reduce analysis time compared to manual analysis of diaphragm electromyography in patients with COPD?
A novel semi-automated method for analyzing diaphragm electromyography during exercise in COPD patients is faster and less prone to inter-rater variability than traditional manual analysis.
May enable reliable EMGdi amplitude analysis without manual QRS selection; leaves open prospective validation in clinical cohorts.
Background: Electrocardiogram (ECG) contamination is present in diaphragm electromyography (EMGdi) recordings. Obtaining EMGdi without ECG contamination is crucial for EMG amplitude analysis. Manually selecting EMGdi in between QRS complexes has been most commonly applied in recent years (manual method). We developed a semi-automated analysis method based on Least Mean Square Adaptive Filtering combined with a synchronously recorded separate ECG channel to remove ECG artifacts from the EMGdi signals. We hypothesized that this approach would shorten analysis duration and might minimize the potential for inter-rater disagreement. Aims: We aimed to evaluate agreement between the semi-automated method and the manual method and inter-rater reliability of the manual method. Methods: EMGdi signals of seven patients with COPD were recorded using an esophageal catheter during an exercise test on a cycle ergometer. Four patients subsequently participated in an inspiratory muscle training (IMT) program for eight weeks. After IMT, the tests were repeated. EMGdi/EMGdiMax as obtained either manually by the two assessors or retrieved from the semi-automated method were compared. Results: Semi-automated EMGdi/EMGdiMax agreed well with values obtained by one of the two manual assessors (assessor 1) both at pre-intervention measurements (mean difference -0.5%, 95% CI -19.6% to 18.6%) and for the pre/post IMT differences (mean difference 1.2%, 95% CI -16.8% to 19.2%). Intra-class correlation coefficients between methods were 0.96 (95% CI: 0.94 to 0.97) at pre-intervention measurements and 0.78 (95% CI: 0.58 to 0.89) for pre/post IMT differences (both p<0.001). EMGdi/EMGdiMax from assessor 2 was systematically lower than from assessor 1 and agreed less well with the semi-automated method both at pre-intervention measurements (mean difference: 9.3%, 95% CI:-11.4% to 29.9%) and for pre/post IMT differences (mean difference 7.0%, 95% CI: -20.4% to 34.4%). Analysis duration of the semi-automated method was significantly shorter (29±9 minutes) than the manual method (82±20 minutes, p<0.001). Conclusion: The developed semi-automated method is more time efficient and will be less prone to inter-rater variability that was observed when applying the manual analysis method. It is, therefore, proposed as a new standard for objective EMGdi amplitude analyses in future studies.
No takes yet. Share an insight, caveat, or question.
Dacha et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: