Novel bipolar RF catheter achieves target ablation zone in ~60 seconds without obstructing renal flow.
Answers are grounded in this paper and cite their sources.
BACKGROUND: Renal denervation with radiofrequency ablation has become an accepted treatment for drug-resistant hypertension. However, there is a continuing need to develop new catheters for high-accuracy, targeted ablation. We therefore developed a radiofrequency bipolar electrode for controlled, targeted ablation through Joule heating induction between 60 and 100 °C. The bipolar design can easily be assembled into a basket catheter for deployment inside the renal artery. METHODS: Finite element modeling was used to determine the optimum catheter design to deliver a minimum ablation zone of 4 mm (W) × 10 mm (L) × 4 mm (H) within 60 s with a 500 kHz, 60 Vp-p signal, and 3 W maximum. The in silico model was validated with in vitro experiments using a thermochromic phantom tissue prepared with polyacrylamide gel and a thermochromic ink additive that permanently changes from pink to magenta when heated over 60 °C. RESULTS: The in vitro ablation zone closely matched the size and shape of the simulated area. The new electrode design directs the current density towards the artery walls and tissue, reducing unwanted blood temperature increases by focusing energy on the ablation zone. In contrast, the basket catheter design does not block renal flow during renal denervation. CONCLUSIONS: This computational model of radiofrequency ablation can be used to estimate renal artery ablation zones for highly targeted renal denervation in patients with resistant hypertension. Furthermore, this innovative catheter has short ablation times and is one of the lowest power requirements of existing designs to perform the ablation.
No takes yet. Share an insight, caveat, or question.
Pérez et al. (2021) studied Drug-resistant hypertension. Bipolar radiofrequency ablation basket catheter for renal denervation vs. Finite element computational model simulation was evaluated on Minimum ablation zone size and shape (target 4 mm × 10 mm × 4 mm at 60–100 °C). A new bipolar radiofrequency ablation catheter achieved an in vitro ablation zone matching the simulated 4×10×4 mm target within 60 s at 3 W max without blocking renal flow.
Synapse has enriched 10 related papers. Consider them for comparative context: