Why the study?
4D flow MRI enables quantitative assessment of helical flow, but normal values and changes across age and gender for helical flow, flow jets, and wall shear stress needed to be generated.
This study provides quantitative normal values for 4D flow MRI parameters in the ascending aorta, demonstrating that wall shear stress and maximum helical forward velocity decrease with age.
Answers are grounded in this paper and cite their sources.
Reference values for helical flow by 4D MRI generated; leaves open clinical utility pending disease-state validation.
Objectives 4D flow MRI enables quantitative assessment of helical flow. We sought to generate normal values and elucidate changes of helical flow (duration, volume, length, velocities and rotational direction) and flow jet (displacement, flow angle) as well as wall shear stress (WSS). Methods We assessed the temporal helical existence ( TH EX ), maximum helical volume ( HV max ), accumulated helical volume ( HV acc ), accumulated helical volume length ( HVL acc ), maximum forward velocity ( maxV for ), maximum circumferential velocity ( maxV circ ), rotational direction ( RD ) and maximum wall shear stress ( WSS ) as reported elsewhere using the software tool Bloodline in 86 healthy volunteers (46 females, mean age 41 ± 13 years). Results WSS decreased by 42.1% and maxV for by 55.7% across age. There was no link between age and gender regarding the other parameters. Conclusion This study provides age-dependent normal values regarding WSS and maxV for and age- and gender-independent normal values regarding TH EX , HV max , HV acc , HVL acc , RD and max V circ . Key Points • 4D flow provides numerous new parameters; therefore, normal values are mandatory. • Wall shear stress decreases over age. • Maximum helical forward velocity decreases over age.
No takes yet. Share an insight, caveat, or question.
Ebel et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: