Does precordial recording of iodinated I 131 human serum albumin dilution curves accurately measure cardiac output compared to arterial blood sampling in normal and heart failure patients?
Does precordial recording of iodinated I 131 human serum albumin dilution curves accurately measure cardiac output compared to arterial blood sampling in normal and heart failure patients?
Precordial recording of iodinated I 131 human serum albumin dilution curves provides a non-invasive alternative to arterial sampling for determining cardiac output with good accuracy.
Answers are grounded in this paper and cite their sources.
May support non-invasive cardiac output measurement; leaves open larger prospective validation before clinical adoption.
Technics for determining the cardiac output by recording the dilution curve of injected iodinated I 131 human serum albumin from precordial sites have been described. By viewing predominantly either the right or left side of the heart and by rapid delivery of the isotope, quantitation of the dilution curves for cardiac output has been possible without critical requirements for placement of the counter. In a series of 26 determinations in normal patients cardiac output values calculated from simultaneously recorded precordial curves and withdrawn arterial blood showed an average agreement within ± 8 per cent. In a smaller series of 8 determinations on patients in heart failure, an average deviation of ± 9 per cent was found.