Why the study?
Does a 30-minute automated office blood pressure measurement agree with daytime ambulatory blood pressure monitoring for diagnosing elevated blood pressure?
Does a 30-minute automated office blood pressure measurement agree with daytime ambulatory blood pressure monitoring for diagnosing elevated blood pressure?
A 30-minute automated office blood pressure measurement protocol provides mean readings comparable to daytime ambulatory monitoring and can effectively classify white-coat and masked hypertension.
Answers are grounded in this paper and cite their sources.
Supports 30-min OBPM as a feasible office option; leaves open its reliability for individual hypertension classification.
PURPOSE: Current office blood pressure measurement (OBPM) is often not executed according to guidelines and cannot prevent the white-coat effect. Serial, automated, oscillometric OBPM has the potential to overcome both these problems. We therefore developed a 30-minute OBPM method that we compared with daytime ambulatory blood pressure. METHODS: Patients referred to a primary care diagnostic center for 24-hour ambulatory blood pressure monitoring (ABPM) had their blood pressure measured using the same validated ABPM device for both ABPM and 30-minute OBPMs. During 30-minute OBPM, blood pressure was measured automatically every 5 minutes with the patient sitting alone in a quiet room. The mean 30-minute OBPM (based on t = 5 to t = 30 minutes) was compared with mean daytime ABPM using paired t tests and the approach described by Bland and Altman on method comparison. RESULTS: We analyzed data from 84 patients (mean age 57 years; 61% female). Systolic and diastolic blood pressures differed from 0 to 2 mm Hg (95% confidence interval, -2 to 2 mm Hg and from 0 to 3 mm Hg) between mean 30-minute OBPM and daytime ABPM, respectively. The limits of agreement were between -19 and 19 mm Hg for systolic and -10 and 13 mm Hg for diastolic blood pressures. Both 30-minute OBPM and daytime ABPM classified normotension, white-coat hypertension, masked hypertension, and sustained hypertension equally. CONCLUSIONS: The 30-minute OBPM appears to agree well with daytime ABPM and has the potential to detect white-coat and masked hypertension. This finding makes 30-minute OBPM a promising new method to determine blood pressure during diagnosis and follow-up of patients with elevated blood pressures.
No takes yet. Share an insight, caveat, or question.
Wel et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: