Why the study?
How do the outputs of the Technique Laboratories TM-8 and Hatteland BD-100 devices compare for detecting microemboli during cardiopulmonary bypass?
How do the outputs of the Technique Laboratories TM-8 and Hatteland BD-100 devices compare for detecting microemboli during cardiopulmonary bypass?
The Technique Laboratories TM-8 and Hatteland BD-100 microemboli detectors show a linear relationship in output, though variability increases at higher ranges, which is important for comparing published data on cardiopulmonary bypass equipment.
Answers are grounded in this paper and cite their sources.
Outputs may not be interchangeable at high counts; leaves open standardization of microemboli detection across devices in bypass studies.
Detection of gaseous microemboli during cardiopulmonary bypass procedures is important for the clinical evaluation of equipment such as oxygenators and cardiotomy reservoirs. Comparison of published data can be difficult if different detectors are used. Two devices reported in the literature, the Technique Laboratories TM-8 and the Hatteland BD-100, are compared during clinical procedures. The relationship between the outputs of these devices was linear over two ranges, the difference in output amounted to a standard deviation of 11% in the lower range and 38% in the upper range.
No takes yet. Share an insight, caveat, or question.
Clayton et al. (1990) studied this question.