Why the study?
Do different in-line and in vitro blood gas PO2 sensors provide comparable measurements during cardiopulmonary bypass?
Do different in-line and in vitro blood gas PO2 sensors provide comparable measurements during cardiopulmonary bypass?
In-line PO2 sensors demonstrate close agreement with each other during cardiopulmonary bypass, whereas an in vitro blood gas analyzer showed poorer agreement.
Answers are grounded in this paper and cite their sources.
May support interchangeable in-line PO2 sensors during CPB; leaves open standardization with in vitro analyzers.
During 17 cardiopulmonary bypass (CPB) operations, arterial PO2 measurements were taken at 5 min intervals using four different types of PO2 sensor; three in-line (CardioMet-1000, Polytrode, and Gas-Stat) and one blood gas analyser (IL-1302). Comparisons between sensors were made relative to the CardioMet-1000. The three in-line sensors showed close agreement. The mean and standard deviation of the difference between the Gas-Stat and the CardioMet-1000 was 0.7 +/- 4.7 kPa, and between the Polytrode and the CardioMet-1000 was -1.4 +/- 4.0 kPa. Agreement between the IL-1302 blood gas analyser and CardioMet-1000 was not as good, the difference being -7.0 +/- 8.6 kPa.
No takes yet. Share an insight, caveat, or question.
Poslad et al. (1987) studied this question.