Laser-Doppler flowmetry provides a continuous measurement of blood flow without violating the natural state of circulation. The linearity of the laser-Doppler and hydrogen clearance methods of blood flow measurement were compared using a protocol that produced changes in cerebral blood flow that might be experienced in a neurosurgical setting. Cerebral blood flow was measured in both hemispheres of 12 adult cats during the snaring of one common carotid artery, the intracisternal injection of 5 mg of 5-hydroxytryptamine creatinine sulfate, and hypervolemic hemodilution, which produced a 25% reduction in blood hematocrit. The percentage of baseline laser-Doppler flowmetry and hydrogen clearance flows showed an acceptable degree of correlation (R2 = 0.762) over the range of cerebral blood flows measured. More rigorous analysis using Bland and Altman's difference against mean test showed that 10 minutes after hemodilution, the two methods displayed a level of variation outside the limits of agreement (-21.85 to 22.03%). Laser-Doppler flowmetry provided a noninvasive and continuous measure of blood flow, increasing the ability to observe instantaneous changes in cerebral microcirculation. However, laser-Doppler flowmetry did not record absolute blood flow, was affected by cerebral tissue shrinkage, and did not accurately measure flow under conditions of changed blood hematocrit.
No takes yet. Share an insight, caveat, or question.
Kramer et al. (1996) studied this question.