Why the study?
Does severe chronic kidney disease increase the risk of ICD related complications in patients receiving an ICD or CRT-D?
Does severe chronic kidney disease increase the risk of ICD related complications in patients receiving an ICD or CRT-D?
Severe chronic kidney disease (GFR <30) is independently associated with an increased risk of pocket hematoma following ICD or CRT-D implantation.
Answers are grounded in this paper and cite their sources.
Severe CKD may heighten hematoma risk after ICD/CRT-D implantation; leaves open effects on other complications and need for targeted strategies.
INTRODUCTION: Implantable cardioverter defibrillator (ICD) implantation has become an accepted therapy for the prevention of sudden cardiac death. However, serious comorbidities such as chronic kidney disease (CKD) are influencing the beneficial effects of ICD therapy. In this study, the association between kidney function and the occurrence of ICD related complications was assessed. METHODS: All patients receiving an ICD or cardiac resynchronization therapy-defibrillator between 1996 and 2012 were included. Renal function was categorized as: glomerular filtration rate (GFR) >90, GFR 30-90 or GFR <30 mL/min/1.73 m(2) . Registered complications were pocket hematoma, pneumothorax, lead complications, and device infection. RESULTS: In 3,147 device recipients, 236 patients (7.5%) suffered from at least 1 complication. Patients with a GFR <30 (n = 110) had a higher event rate for hematoma, pneumothorax, and infection. These patients were older, had a higher incidence of hypertension, diabetes, and a lower body mass index (BMI; P < 0.05). After correcting for these risk factors, hematoma remained independently associated with a GFR <30 mL/min (OR 2.7, CI: 1.05-6.9, P = 0.04). Device infection, pneumothorax, and lead complications were not independently associated with a GFR <30 mL/min/1.73 m(2) . CONCLUSIONS: Patients with CKD suffered from more ICD related complications than patients without kidney disease. This was partially associated with kidney dysfunction itself as was the case with the occurrence of hematoma. However, the high burden of risk factors associated with device complications in patients with renal disease played an important role as well.
No takes yet. Share an insight, caveat, or question.
Buiten et al. (2014) studied this question.