Why the study?
Conventional ECV calculation requires blood hematocrit measurement, whereas synthetic ECV can be estimated without blood sampling, prompting evaluation of their correlation and agreement.
Does synthetic ECV accurately correlate and agree with laboratory ECV in patients undergoing CMR?
Does synthetic ECV accurately correlate and agree with laboratory ECV in patients undergoing CMR?
Synthetic ECV demonstrates excellent correlation and agreement with laboratory ECV, enabling noninvasive quantification of myocardial extracellular space without blood sampling.
Supports noninvasive synthetic ECV in routine CMR; confirms validity across studies for broader adoption.
The calculation of conventional extracellular volume fraction (ECV) requires blood hematocrit (Hct) measurement. Based on the relationship between Hct and blood T1 relaxivity for cardiac magnetic resonance (CMR), a synthetic ECV could be estimated without a blood sampling. The aim of this study was to evaluate the correlation and agreement in the quantification of synthetic ECV and laboratory ECV from conventional Hct measurements. Electronic database searches of PubMed, Web of Science Core Collection, Cochrane advanced search, and EMBASE were performed. The authors employed a meta-analysis using the generic inverse variance method with a random-effects model to estimate the summary correlation coefficient and mean absolute difference between synthetic and laboratory ECV. Of 38 papers, 10 studies comprising 4492 patients were identified. Overall, there was an excellent correlation between synthetic ECV and laboratory ECV (0.95 [95% confidence interval (CI): 0.92 to 0.97]) at 1.5T CMR and (0.91 [95% CI: 0.86 to 0.94]) at 3.0T CMR. The pooled mean difference between synthetic ECV and laboratory ECV was 0.61% (95% CI: 0.23 to 0.98%, I 2 = 0%, p for heterogeneity = 0.67) at 1.5T CMR and 0.24% (95% CI: −0.13 to 0.61%, I 2 = 19%, p for heterogeneity = 0.25) at 3.0T CMR. This study is the first comprehensive systematic review and meta-analysis of synthetic ECV evaluation at CMR. Synthetic ECV demonstrated an excellent correlation with laboratory ECV, with a mean difference of less than 1%, and offers noninvasive and instantaneous quantification of the myocardial extracellular space without the need for blood sampling.
No takes yet. Share an insight, caveat, or question.
Yasuda et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: