This is the authors' abstract. We don't add key points for this paper.
An indexed rotational immobilization system was developed for supine total body irradiation (TBI). Treatment plans had multi-isocentric volumetric modulated arc therapy (VMAT) beams to the upper body and parallel-opposed fields to the lower body, with a 12 Gy prescription dose to >90% of the body and mean lung dose ∼8 Gy. In the end-to-end test, point dose measurements had <10% error. Compared to conventional TBI, the VMAT-based TBI technique increased the mean dose to the body by ∼1.0–1.5 Gy and decreased the mean dose to the lung by ∼1.0–1.5 Gy. Overall treatment time was ∼1.5 h, similar to conventional TBI.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Ouyang et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: