Key result
Permanent pacemaker implantation with a left ventricular lead in the coronary sinus maintained a stable threshold at 1 year in a patient with an inaccessible right ventricle.
Why the study?
Can the coronary sinus be utilized for permanent pacing in patients with an inaccessible right ventricle due to a prosthetic tricuspid valve?
Case Report (n=1)
Can the coronary sinus be utilized for permanent pacing in patients with an inaccessible right ventricle due to a prosthetic tricuspid valve?
Corary sinus pacing is a feasible alternative for permanent pacing in patients with an inaccessible right ventricle due to a prosthetic tricuspid valve.
Answers are grounded in this paper and cite their sources.
May offer pacing alternative in inaccessible right ventricle; hypothesis-generating case report requiring prospective validation.
Unconventional sites are being used for pacing in patients with inaccessible right ventricle like single ventricle, atresia of tricuspid valve and in anomalous venous drainage. Here we report a case in which the right ventricle could not be accessed due to the metallic prosthesis. A 41-year-old lady required triple valve replacement for rheumatic involvement. Permanent pacemaker implantation was done with epicardial lead for bradycardia post operatively. Pulse generator change was needed within 3 years as the pacemaker reached end of life due to high lead threshold. Subsequent pacemaker implantation was done with a left ventricular lead in coronary sinus by percutaneous approach. One year after implantation, the threshold remains stable. Coronary sinus can be utilized for permanent pacing in patients with inaccessible right ventricle due to prosthetic tricuspid valve.
No takes yet. Share an insight, caveat, or question.
Vijayakumar et al. (2013) conducted a case report in Bradycardia with inaccessible right ventricle due to prosthetic tricuspid valve (n=1). Permanent pacemaker implantation with a left ventricular lead in coronary sinus was evaluated on Pacing threshold stability. Permanent pacemaker implantation with a left ventricular lead in the coronary sinus maintained a stable threshold at 1 year in a patient with an inaccessible right ventricle.
Synapse has enriched 10 related papers. Consider them for comparative context: