Key result
Personalized vital sign alarm thresholds adapted from initial monitoring data resulted in fewer alarms than standard age-based thresholds while remaining similar to clinician-chosen thresholds.
Why the study?
Low precision of vital sign alarm systems causes alarm fatigue among clinical staff, leading to increased response times or ignored alarms that compromise inpatient safety.
Do personalized vital sign alarm thresholds reduce the number of alarms compared to age-based thresholds in pediatric inpatients?
Do personalized vital sign alarm thresholds reduce the number of alarms compared to age-based thresholds in pediatric inpatients?
Personalized vital sign alarm thresholds based on initial monitoring data can reduce alarm fatigue in pediatric inpatient settings without compromising the detection of critical events.
Answers are grounded in this paper and cite their sources.
May reduce alarm fatigue in monitored patients; leaves open prospective outcome validation before practice change.
Bedside monitors are intended as a safety net in patient care, but their management in the inpatient setting is a significant patient safety concern. The low precision of vital sign alarm systems leads to clinical staff becoming desensitized to the sound of the alarm, a phenomenon known as alarm fatigue. Alarm fatigue has been shown to increase response time to alarms or result in alarms being ignored altogether and has negative consequences for patient safety. We present methods to establish personalized thresholds for heart rate and respiratory rate alarms. These thresholds are first chosen based on patient characteristics available at the time of admission and are then adapted to incorporate vital signs observed in the first 2 hours of monitoring. We demonstrate that the adapted thresholds are similar to those chosen by clinicians for individual patients and would result in fewer alarms than the currently used age-based thresholds. Personalized vital sign alarm thresholds can help to alleviate the problem of alarm fatigue in an inpatient setting while ensuring that all critical vital signs are detected.
No takes yet. Share an insight, caveat, or question.
Poole et al. (2019) studied Pediatric inpatients. Personalized thresholds for heart rate and respiratory rate alarms vs. Currently used age-based thresholds was evaluated on Number of alarms and similarity to clinician-chosen thresholds. Personalized vital sign alarm thresholds adapted from initial monitoring data resulted in fewer alarms than standard age-based thresholds while remaining similar to clinician-chosen thresholds.