Key result
This editorial critiques a proposed streamlined exercise testing protocol, highlighting methodological concerns regarding VO2 kinetics, work efficiency, and parameter estimation in diverse populations.
Key points are not available for this paper at this time.
Design
Editorial
This viewpoint highlights methodological concerns with a recently proposed streamlined cardiopulmonary exercise testing protocol, emphasizing the need for rigorous validation of VO2 kinetics and peak VO2 measurements.
Rigorous determination of key parameters of oxidative function, namely the maximum O2 uptake ( V ̇ O 2 max V_O₂max ), V ̇ O 2 V_O₂ kinetics and work efficiency, provides invaluable information regarding the O2 transport system and integrated cerebral–pulmonary–cardiovascular–muscular function in health and disease. For instance, these parameters relate mechanistically to lifespan, health span, physical exercise performance and exercise (in)tolerance. Moreover, in patient populations, these parameters can stratify individuals for heart transplant procedures, assess surgical outcomes and present sensitive measures of treatment toxicity (e.g., doxorubicin and anthracyclines in cancer) as well as discriminate therapeutic benefits of imposed countermeasures to disease or age-related debilitation. The ramp-incremental exercise test, with gas exchange measurements, was originally devised to provide information on each of these parameters in a single laboratory visit (Whipp et al., 1981). Indeed, this approach forms the basis of clinical cardiopulmonary exercise testing guidelines worldwide (e.g., Palange et al., 2018). However, over time, it became apparent that V ̇ O 2 V_O₂ kinetics and work efficiency varied considerably with exercise intensity (e.g., Özyener et al., 2001; Rossiter, 2011), and therefore new approaches were needed to evaluate these diverse aspects of oxidative function in a single exercise test. To that end, Santana et al. (2025) are to be applauded for their recognition of the importance of streamlining laboratory testing to facilitate the measurement of multiple oxidative parameters (i.e., V ̇ O 2 peak V_O₂peak , V ̇ O 2 V_O₂ kinetics and, although not discussed therein, work efficiency). However, prior to the implementation of the precise testing protocol advocated by Santana et al. (2025), multiple checks and balances should be considered based upon sound logic and scientific rigour. There needs to be consideration of the on–off asymmetry that normally attends moderate- versus severe-intensity exercise (e.g., Özyener et al., 2001). This unavoidably leads to the very slow overall time constants observed in recovery from V ̇ O 2 peak V_O₂peak , where there is clearly V ̇ O 2 V_O₂ slow component behaviour (e.g., extant in Figure 4b of Santana et al. (2025)). This inaccuracy can be ameliorated by the application of a two-component model for the recovery analyses (Özyener et al., 2001). Use of 30% predicted V ̇ O 2 V_O₂ reserve to determine moderate-intensity V ̇ O 2 V_O₂ kinetics is infeasible for patient populations with moderate to severe impairments in O2 transport and/or utilization. For example, patients with heart failure or chronic obstructive pulmonary disease can have V ̇ O 2 peak V_O₂peak values that are less than 30% predicted V ̇ O 2 V_O₂ reserve. Also, in severe heart failure, even unloaded pedalling can be attended by a raised arterial lactate and is therefore no longer in the moderate-intensity domain, complicating kinetics analysis (Özyener et al., 2001). While the participants in Santana et al.’s study fell within the range of ‘normal’ aerobic fitness (based on group mean values, males achieved 100% and females achieved 106% of predicted V ̇ O 2 peak V_O₂peak ), their kinetics were measured at 36–37% of predicted V ̇ O 2 peak V_O₂peak . These V ̇ O 2 V_O₂ values could easily exceed the gas exchange threshold (GET) in a deconditioned population (a guideline value to identify deconditioning is GET < 40%). Therefore, consideration of the time constant for moderate-intensity V ̇ O 2 V_O₂ kinetics using this protocol also requires confirmation that the participant is exercising below GET. Repeated transitions are requisite for accurate kinetics parameter analysis from breath-by-breath data (e.g., Benson et al., 2017). Such procedures would likely have removed the poor fit and systematic bias evident in Figure 4. The V ̇ O 2 peak V_O₂peak protocol should ensure that V ̇ O 2 max V_O₂max is validated and therefore measured accurately (Poole & Jones, 2017). Data need to be carefully screened for aberrant behaviour. Specifically, that V ̇ O 2 V_O₂ remains at peak for 40 s into ‘recovery’ in Figure 2 stands in opposition to a legion of prior observations and mechanistic underpinnings of exercise energetics (e.g., Özyener et al., 2001). Finally, Santana et al. report that V ̇ O 2 peak V_O₂peak is much higher in males (per kg) than females, and yet their V ̇ O 2 V_O₂ on-kinetics is far slower. This does not cohere with the compelling weight of published evidence to the contrary (see Poole & Jones, 2012). Attention to these and additional considerations will lessen the ‘streamlined’ nature of the testing protocol proposed by Santana et al. (2025) but is essential for ensuring accurate parameter estimation and to enhance translation of their approach to patients and other defined populations. A fundamental tenet of science is that it be self-correcting with expeditious rectification limiting pernicious sequelae. Today we are privileged to exist in a space and time where, as in this instance, science can be corrected in real time. We hope that such an exercise, as demonstrated by this Viewpoint, will be of interest and pertinent to scientists at all stages of their careers and help reinforce confidence that science is indeed self-correcting. Both authors have read and approved the final version of this manuscript and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. All persons designated as authors qualify for authorship, and all those who qualify for authorship are listed. Harry Rossiter is a visiting Professor at the University of Leeds, UK. No funding was received for this work.
No takes yet. Share an insight, caveat, or question.
Poole et al. (2025) reported an editorial. This editorial critiques a proposed streamlined exercise testing protocol, highlighting methodological concerns regarding VO2 kinetics, work efficiency, and parameter estimation in diverse populations.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: