Key result
Longer lead dwell time was significantly associated with fibrotic reactions within the walls of the heart and veins (OR 1.111, p<0.001) in patients with cardiac implantable electronic devices.
Why the study?
The extent of fibrous connective tissue accumulation around intracardiac leads visible on TEE, and its predisposing factors, are not well defined.
Observational (n=936)
No
Odds Ratio: 1.111
p-value: p=<0.001
Lead dwell time and excessive lead looping are major predisposing factors for lead-associated echocardiographic abnormalities such as fibrotic reactions and tricuspid dysfunction in patients with CIEDs.
May inform extraction risk stratification; leaves open whether earlier removal alters fibrosis progression.
BACKGROUND: The constant interaction between intracardiac leads and the heart and veins results in excessive accumulation of fibrous connective tissue around the leads. The extent of this pathological phenomenon, which is visible on transesophageal echocardiography (TEE), and predisposing factors are not well defined. METHODS: We examined 936 transesophageal echocardiograms prior to transvenous lead extraction (TLE) performed at a high-volume centre between 2015 and 2019. RESULTS: The most important echocardiographic findings were fibrous binding sites between leads and cardiovascular structures, lead-to-lead adhesions, excessive lead loops, lead-dependent tricuspid dysfunction (LDTD), asymptomatic masses on endocardial leads (AMEL) and vegetations. Fibrotic reaction within the walls of the heart and veins correlated with the presence of lead loops (OR = 1.771; p < .01) and lead dwell time (OR = 1.111; p < .001). Women were more likely to have excessive lead loops (OR = 1.639; p < .01), and the occurrence of loops increase with the number of implanted leads (OR = 2.557; p < .001). Heart failure (OR = 4.016; p < .001), lead looping (OR = 2.603; p < .01) and longer cumulative lead dwell time (OR = 1.017; p < .05) increased the likelihood of LDTD. A variety of AMEL were identified in this study, most commonly in patients with older leads (OR = 1.043; p < .001). CONCLUSIONS: Lead dwell time is the main factor predisposing to the occurrence of most lead-associated phenomena visualized by TEE in patients with cardiac implantable electronic devices (CIED). Excessive looping of the lead is an important cause of fibrous binding sites and LDTD. AMEL are frequently detected in CIED patients, and their various forms concurrent with vegetations could represent an evolutionary stage of lead-associated masses.
No takes yet. Share an insight, caveat, or question.
Nowosielecka et al. (2020) conducted an observational in Cardiac implantable electronic devices (CIED) requiring transvenous lead extraction (n=936). Lead dwell time vs. Shorter lead dwell time was evaluated on Fibrotic reaction within the walls of the heart and veins (OR 1.111, p=<0.001). Longer lead dwell time was significantly associated with fibrotic reactions within the walls of the heart and veins (OR 1.111, p<0.001) in patients with cardiac implantable electronic devices.