Key result
A newly developed UHPLC-MS/MS method for quantifying apixaban, dabigatran, edoxaban, and rivaroxaban in human serum fulfilled validation criteria and was successfully applied to 345 routine patient samples.
A newly developed and validated UHPLC-MS/MS method can accurately quantify four major DOACs in human serum, though the observed instability of edoxaban at 30°C may complicate routine sample transport.
BACKGROUND: Direct oral anticoagulants (DOACs) are prescribed for anticoagulation in patients with atrial fibrillation and venous thromboembolic disease. Fixed doses are recommended, but measuring their serum drug concentrations as a basis for dose adjustments may be useful in some clinical settings. METHODS: An ultra-high-performance liquid chromatography-tandem mass spectrometry method for the analysis of the DOACs apixaban, dabigatran, edoxaban, and rivaroxaban in human serum was developed and validated. A 100-µL serum sample was handled using a pipetting robot. Protein precipitation was performed with 375 µL of 1% formic acid in acetonitrile (vol/vol), and phospholipid removal was performed using a Waters Ostro 96-well plate. The injection volume was 1 µL, and run time was 3.0 minutes. RESULTS: The calibration range was 5-800 nmol/L. The between-day precision relative SDs were in the range of 3.3%-10%. Recoveries ranged from 85% to 105%, and matrix effects from 88% to 102%, when corrected with internal standard. Edoxaban was, in contrast to the other DOACs, unstable when stored for more than 6 hours at 30°C. The suitability of the method was demonstrated by analyzing routine samples from 345 patients undergoing anticoagulation treatment. CONCLUSIONS: The developed method fulfilled the set validation criteria, and its suitability was demonstrated in a routine setting. The instability of edoxaban may complicate the transport of routine samples to the laboratory.
No takes yet. Share an insight, caveat, or question.
Lindahl et al. (2018) studied Patients undergoing anticoagulation treatment with DOACs (n=345). UHPLC-MS/MS method for DOAC quantification was evaluated on Method validation parameters (precision, recovery, matrix effects) and clinical application suitability. A newly developed UHPLC-MS/MS method for quantifying apixaban, dabigatran, edoxaban, and rivaroxaban in human serum fulfilled validation criteria and was successfully applied to 345 routine patient samples.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: