Key result
An abbreviated automated office blood pressure measurement protocol using the mean of two measurements showed minimal differences (-0.1 mmHg) and no significant hypertensive reclassifications compared to the standard three-measurement protocol.
Why the study?
Unattended automated office blood pressure measurement is observer-independent but time and space consuming, prompting evaluation of whether the protocol can be abbreviated without losing accuracy.
Does an abbreviated AOBPM protocol (2 measurements or 1 measurement) maintain accuracy compared to a standard 3-measurement protocol in patients undergoing blood pressure measurement?
Observational (n=210)
No
Does an abbreviated AOBPM protocol (2 measurements or 1 measurement) maintain accuracy compared to a standard 3-measurement protocol in patients undergoing blood pressure measurement?
An abbreviated AOBPM protocol using the mean of two measurements is accurate and comparable to a standard three-measurement protocol, potentially saving time and resources in clinical practice.
Answers are grounded in this paper and cite their sources.
Abbreviated 2-measurement AOBPM shows minimal differences; leaves open prospective validation before protocol changes.
BACKGROUND: Blood pressure measurement (BPM) is one of the most often performed procedures in clinical practice, but especially office BPM is prone to errors. Unattended automated office BPM (AOBPM) is somewhat standardised and observer-independent, but time and space consuming. We aimed to assess whether an AOBPM protocol can be abbreviated without losing accuracy. DESIGN: In our retrospective single centre study, we used all AOBPM (AOBPM protocol of the SPRINT study), collected over 14 months. Three sequential BPM (after 5 minutes of rest, spaced 2 minutes) were automatically recorded with the patient alone in a quiet room resulting in three systolic and diastolic values. We compared the mean of all three (RefProt) with the mean of the first two (ShortProtA) and the single first BPM (ShortProtB). RESULTS: We analysed 413 AOBPM sets from 210 patients. Mean age was 52±16 years. Mean values for RefProt were 128.3/81.3 mmHg, for ShortProtA 128.4/81.4 mmHg, for ShortProtB 128.8/81.4 mmHg. Mean difference and limits of agreement for RefProt vs. ShortProtA and ShortProtB were -0.1±4.2/-0.1±2.8 mmHg and -0.5±8.1/-0.1±5.3 mmHg, respectively. With ShortProtA, 83% of systolic and 92% of diastolic measurements were within 2 mmHg from RefProt (67/82% for ShortProtB). ShortProtA or ShortProtB led to no significant hypertensive reclassifications in comparison to RefProt (p-values 0.774/1.000/1.000/0.556). CONCLUSION: Based on our results differences between the RefProt and ShortProtA are minimal and within acceptable limits of agreement. Therefore, the automated procedure may be shorted from 3 to 2 measurements, but a single measurement is insufficient.
No takes yet. Share an insight, caveat, or question.
Vischer et al. (2021) conducted an observational in Arterial hypertension (n=210). Abbreviated AOBPM protocol (mean of first two measurements) vs. Reference AOBPM protocol (mean of three measurements) was evaluated on Mean difference in systolic blood pressure between reference protocol and abbreviated protocol A (MD -0.1, p=0.203). An abbreviated automated office blood pressure measurement protocol using the mean of two measurements showed minimal differences (-0.1 mmHg) and no significant hypertensive reclassifications compared to the standard three-measurement protocol.