Why the study?
Evidence regarding contemporary implantable cardioverter-defibrillator (ICD) battery longevity in a real-world setting was lacking.
Does ICD manufacturer affect battery longevity in patients undergoing ICD implantation?
Does ICD manufacturer affect battery longevity in patients undergoing ICD implantation?
Real-world ICD battery longevity matches industry projections at 5 years, but significant differences exist between manufacturers, with Medtronic and Biotronik devices showing earlier depletion.
Real-world ICD longevity data inform replacement planning; leaves open manufacturer-specific durability comparisons.
BACKGROUND: There is a lack of evidence regarding contemporary implantable cardioverter-defibrillator (ICD) battery longevity. Our aim was to assess battery longevity in ICDs in a real-world setting. METHODS: Retrospective cross-sectional single center study of a prospectively collected database of consecutive patients who underwent ICD implantation from January 2010 to December 2015. Clinical data and battery longevity of all manufacturers were collected. RESULTS: A total of 351 patients (84.6% males, mean age of 61 ± 12 years) were included in the study (292 VVI; 6 VDD; 53 DDD). All manufacturers (Abbott, Biotronik, Boston, Medtronic and Microport) were equally represented in the study (p = 0.110). Median battery longevity was 10.8 years (11 years for VVI and 8.5 for DDD). After a follow-up time of 5 years, 98% of VVI and DDD were still in service (vs. industry-projected longevity of 98%). During this time, 89 patients (25.4%) underwent device replacement - 69 patients (77.5%) due to battery depletion, 6 patients due to infection, 3 patients due to dysfunction and 13 patients due to upgrade to CRT-D. Patients with Medtronic or Biotronik ICDs had a greater probability of being replaced earlier due to battery depletion (Biotronik HR 6.87, 95% CI 2.54-18.58, p < 0.001; Medtronic HR 6.08, 95% CI 2.45-15.06 p < 0.001). CONCLUSIONS: VVI and DDD ICD battery longevity matched industry-projected longevity after 5 years of follow-up. Medtronic and Biotronik ICDs appeared to have an earlier battery depletion. Further randomized studies are required to ensure optimal care.
No takes yet. Share an insight, caveat, or question.
Fernandes et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: