The Medtronic Sprint Fidelis ICD lead is associated with frequent abnormal R-wave sensing and a 2.8% lead revision rate in the first year, highlighting potential safety issues with downsized ICD leads.
Answers are grounded in this paper and cite their sources.
May prompt closer R-wave monitoring with downsized leads; leaves open long-term failure risk.
INTRODUCTION: The diameter of implantable cardioverter-defibrillator (ICD) leads has become progressively smaller over time. However, the long-term performance characteristics of these smaller ICD leads are unknown. METHODS: We retrospectively evaluated 357 patients who underwent implantation of a Medtronic Sprint Fidelis defibrillating lead at two separate centers between September 2004 and October 2006. Lead characteristics were measured at implant, at early follow-up (1-4 days post implant), and every 3-6 months thereafter. RESULTS: During the study period, 357 patients underwent implantation of the Medtronic Sprint Fidelis lead. The mean R-wave measured at implant through the device was not different (P = NS) when compared with that measured at first follow-up (10.5 +/- 5.0 mV vs 10.7 +/- 5.1 mV). Forty-one patients (13%) had an R-wave amplitude <or= 5 mV measured through the device at implant. Of those patients with an R-wave amplitude <or= 5 mV at implant measured through the device, 63% (n = 26) remained <or= 5 mV for the duration of follow-up. The mean time to R-wave amplitude <or= 5 mV was 96.2 +/- 123 days. During follow-up, 65 (18%) patients developed R-wave <or= 5 mV. Overall 10 lead revisions (2.8%) were performed during the first year of follow-up. CONCLUSION: Abnormal R-wave sensing is frequently observed during follow-up with the Medtronic Fidelis ICD lead. Lead revision was necessary in 2.8% of the patients, most often (8 of 10) due to abnormal R-wave sensing along with elevated pacing threshold. Whether this issue is limited to this lead or reflects a potential problem with all downsized ICD leads merits further investigation.
No takes yet. Share an insight, caveat, or question.
Kenigsberg et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: