Why the study?
Heart rate variability devices differ in their measurement methods, including timing, pulse versus R waves, and signal processing, leaving the level of agreement between systems unclear.
Does the Polar Nightly Recharge function agree with the Polar H10 chest strap and Kubios app in measuring weekly heart rate variability changes in recreational athletes?
Does the Polar Nightly Recharge function agree with the Polar H10 chest strap and Kubios app in measuring weekly heart rate variability changes in recreational athletes?
Despite good correlation and reliability between the Polar Nightly Recharge and Polar H10 chest strap with Kubios app, there is significant discordance in the interpretation of weekly HRV changes, suggesting they should not be used interchangeably.
Answers are grounded in this paper and cite their sources.
Discordance in weekly HRV change interpretation across devices warrants caution in athletes; leaves open need for standardized validation before broader use.
Monitoring heart rate variability has been commonly performed by different devices which differ in their methods (i.e., night recording vs. upon awakening measure, pulse vs. R waves, and software signal processing), Thus, the purpose of this study was to determine the level of agreement between different methods of heart rate variability monitoring, represented in two different systems (i.e., the Polar Nightly Recharge™ function present in Polar sport watches and the Polar H10 chest strap synchronized with the Kubios app). A group of 11 recreational athletes performed a concurrent training program for eight weeks and heart rate variability was daily monitored through both devices. Very large correlation ( r = 0.714) and good reliability (ICC = 0.817) were obtained between devices through the entire training program. The magnitude-based inference method was also applied to determine the likelihood of the change concerning the smallest worthwhile change. From a baseline corresponding to the first two weeks of the training program, the weekly heart rate variability changes of the following six weeks were determined for each participant with each device. Despite the large correlation and good reliability between devices, there was a 60.6% of discordance in the likelihood interpretation of the change for the 66 weeks evaluated, explained by the random errors found. Thus, practitioners should be aware of these differences if their training groups use different devices or if an athlete interchanges them. The different nuances of each device can condition the heart rate variability data variation which could compromise the interpretation of the autonomic nervous system modulation.
No takes yet. Share an insight, caveat, or question.
Ruiz‐Alias et al. (2022) studied this question.