Why the study?
To evaluate in real time during surgery whether a previously developed machine-learning algorithm can detect introduced arterial pressure transducer drop errors from waveform features alone.
Does a machine-learning trained algorithm accurately detect arterial pressure transducer drops in real-time among surgical patients undergoing invasive hemodynamic monitoring?
Does a machine-learning trained algorithm accurately detect arterial pressure transducer drops in real-time among surgical patients undergoing invasive hemodynamic monitoring?
A machine-learning algorithm can accurately detect arterial pressure transducer drops of ≥10 mmHg in real-time with a very low false-positive rate, potentially improving the safety of invasive hemodynamic monitoring.
Supports real-time transducer error detection during arterial monitoring; leaves open whether adoption improves hemodynamic management or outcomes.
Invasive arterial pressure monitoring is frequently used in operating room and intensive care unit patients to provide continuous assessment of hemodynamics. We have previously developed an algorithm trained to detect error of >10 mmHg in leveling of the arterial pressure transducer from only the features of the waveform itself. In the present study we used the detection algorithm in real-time during surgical cases where the primary team had placed an arterial line for hemodynamic monitoring and tested its ability to detected introduced transducer drop errors. Patients having surgery at UCI Medical Center who were 18 years or older and were expected to have invasive arterial pressure monitoring as part of their anesthetic care were eligible for enrollment. The arterial pressure waveform was captured during surgery and at random times the transducer was dropped from its secured location. The ability of the previously trained algorithm to detect these drops from the waveform alone was assessed in real-time as well as false-positives. A total of 50 patients were ultimately included for over 80 h of monitoring time. A total of 78 drop tests were performed in that time. Overall, the algorithm detected 69 of these drops (88%), however detection for errors of 10 mmHg or more was 98% and for >12 mmHg was 100%. The overall false positive rate was 0.65% and balanced accuracy was 0.99 for errors >10 mmHg. The algorithm was highly sensitive for detection of transducer drops resulting in error of ≥10 mmHg in real-time, with a very low false-positive rate. This algorithm may be of value in applications utilizing the arterial line for input where safety is essential.
No takes yet. Share an insight, caveat, or question.
Rinehart et al. (2022) studied this question.