The modified binning algorithm in newer St Jude Medical ICDs may inappropriately inhibit defibrillation during polymorphic ventricular tachycardia, leading to fatal outcomes.
A patient with ischaemic cardiomyopathy was found dead 3 months after the implantation of a primary prevention biventricular cardioverter-defibrillator. The last event on his device showed correct detection of ventricular tachycardia (VT) in the VT-2 zone. However, defibrillation was inhibited during redetection because a sequence of unbinned events (marked with dashes) and intervals that fell in the VT-1 monitored zone (marked as ‘VS’), resulted in a ‘Return to Sinus’ response ending the episode. Newer generations of St Jude Medical devices use a modified binning algorithm when a monitor zone is programmed (Monitor Zone Revision 3) such that events are ‘unbinned’ if the current interval falls into a different zone as the average of the last four beats (Interval Average), unlike the standard binning schema. This may lead to numerous unbinned events during polymorphic VT. Hence, only regular monomorphic VT or very fast ventricular fibrillation episodes that do not straddle tachycardia zones will result in consistent detection and delivery of therapy. The revised binning algorithm has not been tested in a clinical study and has only been bench tested on historical cases. This case highlights the importance of thoroughly understanding the increasingly complex behaviour of modern implantable cardioverter-defibrillators. The full-length version of this report can be viewed at: http://www.escardio.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: