PURPOSE: To report clinical and spectrographic analyses of 2 explanted hydrophilic acrylic intraocular lenses (IOLs). SETTING: John A. Moran Eye Center, Salt Lake City, Utah, USA, and Johannes Gutenberg-University, Department of Ophthalmology, Mainz, Germany. METHODS: We report 6 cases of opacification of MemoryLens IOLs (Ciba Vision) that occurred approximately 2 years after implantation. The anterior and posterior surfaces of the IOLs had a white, frosted appearance, and the IOLs' interior looked brown, similar to the appearance of a brunescent cataract. Two of the IOLs were explanted because of a significant decrease in visual acuity. The IOLs were sent for further clinicopathologic analysis including scanning electron microscopy and energy dispersive x-ray spectroscopy (EDX). RESULTS: Microscopic analysis revealed multiple, fine, granular deposits of various sizes on the surface of the lens optics. The EDX analysis showed the presence of calcium within the deposits. CONCLUSIONS: Our cases show that hydrophilic acrylic IOLs can opacify even years after implantation. Analysis of the explanted IOLs revealed calcification.
No takes yet. Share an insight, caveat, or question.
Tehrani et al. (2004) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: