Key result
The developed wearable IoT pulse oximeter demonstrated high accuracy, with an average measurement difference of ± 2.8 BPM for heart rate and ± 1.5% for oxygen saturation compared to a standard clinical monitor.
A newly developed low-cost, wearable IoT pulse oximeter demonstrated high accuracy comparable to standard medical devices, enabling real-time remote monitoring.
Answers are grounded in this paper and cite their sources.
Promising accuracy supports wearable IoT oximeter development; leaves open prospective clinical validation before adoption.
This paper discusses a device for measuring oxygen saturation (SpO2) and heart rate as parameters of the representations of heart conditions. SpO2 device that have been made has a small dimension, wearable and high mobility with battery as the main power source. The device connects to a node MCU as a data processor and an internet network gateway to support internet of things applications. Data sent to the Internet cloud can be accessed online and real time via website for further analysis. The error rate at heart rate measurement is ± 2.8 BPM and for oxygen saturation (SpO2) is ± 1.5%. Testing data transmission delay until it can be displayed on website is 3 second that depends on internet traffic conditions.
No takes yet. Share an insight, caveat, or question.
Adiputra et al. (2018) studied this question. Wearable IoT SpO2 and heart rate device vs. Mindray patient monitor was evaluated on Measurement error rate for heart rate and SpO2 (± 2.8 BPM (heart rate), ± 1.5% (SpO2)). The developed wearable IoT pulse oximeter demonstrated high accuracy, with an average measurement difference of ± 2.8 BPM for heart rate and ± 1.5% for oxygen saturation compared to a standard clinical monitor.
Synapse has enriched 10 related papers. Consider them for comparative context: