Key result
Amiodarone treatment for one year resulted in a severe increase in atrial pacing threshold and loss of atrial capture, which normalized upon drug discontinuation.
Case Report (n=1)
Amiodarone may cause a clinically significant, reversible, and atrium-specific increase in pacing thresholds leading to loss of capture.
Answers are grounded in this paper and cite their sources.
Amiodarone may cause reversible atrial capture loss in paced patients; single case leaves open need for prospective confirmation before practice change.
Increases of pacing threshold stimulation are well documented with different antiarrhythmic drugs, but not with amiodarone. We report a case of a patient with dual-chamber pacing, with stable thresholds on repeated measurements on the last year, who presented severe increase of atrial pacing threshold resulting in loss of atrial capture after a year of treatment with amiodarone. Thresholds were normalized once amiodarone was removed. Ventricular thresholds were not affected.
No takes yet. Share an insight, caveat, or question.
Grande et al. (2011) conducted a case report in Dual-chamber pacing (n=1). Amiodarone was evaluated on Atrial pacing threshold. Amiodarone treatment for one year resulted in a severe increase in atrial pacing threshold and loss of atrial capture, which normalized upon drug discontinuation.