Why the study?
Does biplane transesophageal echocardiography (B-TEE) guidance improve the feasibility and safety of transseptal puncture and left atrial cannula introduction?
Does biplane transesophageal echocardiography (B-TEE) guidance improve the feasibility and safety of transseptal puncture and left atrial cannula introduction?
Biplane transesophageal echocardiography provides reliable guidance for transseptal puncture and left atrial cannula placement, potentially allowing for safe percutaneous left heart bypass without radiographic guidance.
Answers are grounded in this paper and cite their sources.
B-TEE may aid transseptal needle visualization; hypothesis-generating for fluoroscopy-free left atrial access in larger cohorts.
For introduction of a left atrial (LA) cannula by the transseptal puncture technique, we examined the feasibility of using biplane transesophageal echocardiography (B-TEE). A transseptal puncture was performed on 15 patients (3 male; 12 female; mean age, 48.9 +/- 11.2 years) by B-TEE guide during percutaneous transvenous mitral comissurotomy (PTMC). The entire Brocken-brough needle and the position of its tip were clearly observed in the right atrium by a longitudinal image of B-TEE in all patients (100%), and in 2 (13%) of them also by the transverse image of B-TEE. The puncture was about 1 cm caudal from the center of the fossa ovalis to avoid any large residual atrial septal defect. After transseptal puncture, a Mullin's sheath (7 Fr) and a dilator (14 Fr) were inserted into the left atrium in order, and then an Inoue's balloon catheter (12 Fr) was introduced without difficulty into the left atrium in all patients. With contrast injection, the position of the sheaths's tip was clearly confirmed by B-TEE. Left heart bypass support (left atrial-femoral artery bypass or AAB) was performed on 2 patients after percutaneous introduction of the LA cannula using this technique, and both were successfully supported and survived. With B-TEE guidance, the Brockenbrough atrial septal puncture and introduction of the LA cannula into left atrium can be performed easily. Thus, percutaneous left heart bypass can be set up quickly and safely even in an intensive care unit or outpatient emergency room without radiographic guidance.
No takes yet. Share an insight, caveat, or question.
Kyo et al. (1992) studied this question.