Key result
Repeat transvenous lead extraction shows no difference in long-term mortality compared to single extraction.
Why the study?
Data regarding repeat transvenous lead extraction are scarce.
Does repeat transvenous lead extraction affect procedural safety, efficacy, and long-term mortality compared to a single extraction in patients with cardiovascular implantable electronic devices?
Cohort (n=3,654)
Yes
Does repeat transvenous lead extraction affect procedural safety, efficacy, and long-term mortality compared to a single extraction in patients with cardiovascular implantable electronic devices?
Answers are grounded in this paper and cite their sources.
Repeat transvenous lead extraction occurs in 5.28% of TLE cases—primarily driven by lead failure and the retention of active or abandoned leads during initial extraction—and carries similar procedural success and long-term survival as initial extraction.
Background: Data regarding repeat transvenous lead extraction (TLE) are scarce. The aim of study was to explore the frequency of repeat TLE, its safety, predisposing factors, as well as effectiveness of repeat procedures. Methods: Retrospective analysis of a large single-center database of 3654 TLEs. Results: Repeat TLE was a rare occurrence (193, i.e., 5,28% among 3654 TLEs). Subsequent re-extractions occurred in 12.21% of the patients. Lead failure was the most common cause of re-extraction (51.16%). Cox regression analysis showed that patients who were older at first implantation [HR = 0.987; p = 0.003], had infection-related TLE [HR = 0.392; p < 0.001] and complete procedural success [HR = 0.544; p = 0.034] were less likely to undergo repeat TLE. Functional leads left in place for continuous use [HR = 1.405; p = 0.012] or superfluous leads left in place (abandoned) [HR = 2.370; p = 0.011] were associated with an increased risk of undergoing a repeat procedure. Overall mortality in patients with repeat TLE and subsequent re-extraction in the entire FU period was similar to that in patients without a history of re-extraction [HR = 0.949; p = 0.480]. Conclusions: Repeat TLE was a rare occurrence (5.28%) among TLEs. Left of both active and nonactive leads during TLE increased the risk of re-extraction. Re-extraction has no effect on the long-term mortality.
No takes yet. Share an insight, caveat, or question.
Kutarski et al. (2022) conducted a cohort in Cardiovascular implantable electronic device (CIED) recipients undergoing transvenous lead extraction (TLE) (n=3,654). Repeat transvenous lead extraction (re-extraction) vs. Single transvenous lead extraction (no history of re-extraction) was evaluated on Overall long-term mortality (HR 0.949, 95% CI 0.822-1.097, p=0.480). Repeat transvenous lead extraction occurred in 5.28% of procedures and had no significant effect on long-term mortality compared to a single extraction (HR 0.949, p=0.480).