Answers are grounded in this paper and cite their sources.
In this paper the examples of three-component dynamometers manufactured by using the structure of parallel beams are described. And the deformation and the strain distribution of the onecomponent cantilever dynamometer that is the fundamental shape of the parallel-beam dynamometer are analyzed by Finite Element Method. The main results are as follows: (1) If strain gauges are put on the boundary points between the parallel-beam part and the supporting part, the strain output is the largest. The distribution of normal strain in parallel-beam region is in a straight line. (2)If a Wheetstone bridges circuit is constituted of gauges put on those points symmetrically, the cross sensitivity comes to be extinct. (3) The deflection of the loading point is very small and the strain output per unit displacement is larger than that of a ring dynamometer. (4) The parallel-beam dynamometer is designed easily and a three-component dynamometer can be made compactly.
No takes yet. Share an insight, caveat, or question.
Tani et al. (1982) studied this question.