Why the study?
Does noninvasive monitoring with a multiprogrammable pacing system allow assessment and compensation for early post-implant changes in pulse width threshold in patients following pacemaker implantation?
Does noninvasive monitoring with a multiprogrammable pacing system allow assessment and compensation for early post-implant changes in pulse width threshold in patients following pacemaker implantation?
Noninvasive monitoring with a multiprogrammable pacing system safely and accurately detects early post-implant changes in pulse width threshold, allowing for appropriate reprogramming and potentially improving pacemaker longevity.
Answers are grounded in this paper and cite their sources.
Early PWT rises may warrant close post-implant monitoring and adjustments; leaves open effects on longevity in larger cohorts.
Twenty-two patients were studied for 8 to 12 weeks following pacemaker implantation. A unipolar multiprogrammable pacing system was used to assess and compensate for early post-implant changes in pulse width threshold (PWT). Changes in PWT were estimated by noninvasive measurements of the lowest pulse width required for 100% capture at constant voltage. All patients showed an early increase in PWT. The magnitude of this change was unpredictable and ranged from 2- to 10-fold. For pacing, we selected an appropriate margin of safety to be a pulse width three times the PWT. During the study period, 4 patients showed critical increases in PWT, requiring reprogramming to an increased pulse width as above, whereas only one patient required lead repositioning. The majority of patients achieved a stable low PWT and were reprogrammed to a reduced pulse width. The programmable pacing system provides a safe accurate technique for monitoring early changes in PWT, defines the need for appropriate alterations in programmable functions, decreases the risk of early pacemaker failure, and offers promise for improved pacemaker longevity in many patients.
No takes yet. Share an insight, caveat, or question.
Scoblionko et al. (1981) studied this question.