Key result
An inappropriate ICD shock caused by ventricular lead dislodgement induced ventricular fibrillation, resulting in severe hypoxic brain damage and death 1 week later.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Inappropriate ICD shocks due to lead dislodgement can induce fatal ventricular fibrillation, underscoring the critical need for immediate device interrogation and deactivation when lead dislodgement is suspected.
INTRODUCTION: Inappropriate implantable cardioverter defibrillator (ICD) therapy carries a low but relevant risk of ventricular proarrhythmia. In the present case, the extremely rare event of a fatal arrhythmia caused by inappropriate therapy is reported. Dislodgement of the ventricular lead to the level of the tricuspid annulus led to additional sensing of the atrial signal during sinus tachycardia. Spuriously, ventricular fibrillation was sensed and induced inappropriate ICD shocks. The fourth inappropriate shock caused ventricular fibrillation, which was subsequently undersensed by the dislodged lead due to low ventricular amplitudes. The ICD started antibradycardic pacing during ventricular fibrillation. After initial successful resuscitation, the patient died 1 week later due to severe hypoxic brain damage. Although not preventable in the present case, it underlines the necessity of immediate interrogation of the ICD after ICD therapy and deactivation of the ICD in the setting of a dislodged endocardial lead and intensive care monitoring of the patient until revision.
No takes yet. Share an insight, caveat, or question.
Veltmann et al. (2006) conducted a case report in Inappropriate ICD shock (n=1). Inappropriate ICD shock was evaluated on Death. An inappropriate ICD shock caused by ventricular lead dislodgement induced ventricular fibrillation, resulting in severe hypoxic brain damage and death 1 week later.
Synapse has enriched 10 related papers. Consider them for comparative context: